18 19
Cassette Seal SCS001
Compressor Filter SCF002
Biological Filter SCF032
PART RPI PART # STERILIZER PM TIP
Refer to the
“Installation and Monthly Maintenance Guide”
included with the
Cassette Seal, or see page 22.
Remove the old Cassette Seal. Clean the cassette groove and inspect the
cassette tray and lid for damage. Lubricate and install the new Cassette Seal.
The OEM recommends lubrication of Cassette Seal every 30 days and
replacement every 6 months or 500 cycles – whichever comes first. Clean the
outer edge of cassette tray and the exposed area of the Cassette Seal weekly.
Apply Sci-Dry™ (RPI Part #SCA054) every 10 cycles and after cleaning.
Remove the pump filter cover. Remove filter and clean out any dust or debris.
Replace filter and cover. If you find that the old Air Compressor Filter is wet
or shows signs of water staining, then the check valve is leaking and must be
replaced to protect the air compressor from damage. RPI offers a Check Valve
Kit (RPI Part #SCK011) which includes translucent tubing making it easy to see
a leaky check valve.
The OEM recommends replacing the Air Compressor Filter every 6 months or
500 cycles – whichever comes first.
This Biological Filter fits STATIM 2000 models with serial numbers
2101DK0274 and up, or if the autoclave has been upgraded. (Note: If your
autoclave is not equipped with this filter set-up, then you will not need this
filter).
Note the flow arrow direction on the filter body (flow direction is from the air
compressor to the check valve). If you find that the old Biological Filter is wet
or shows signs of water staining, then the check valve is leaking and must be
replaced to protect the air compressor from damage. RPI offers a Check Valve
Kit (RPI Part #SCK011) which includes translucent tubing making it easy to see
a leaky check valve.
The OEM recommends replacing the Biological Filter every 6 months or 500
cycles – whichever comes first.
4.0 - PLANNED MAINTENANCE (PM) CHECKLIST
Sterilizer PM Kit (RPI Part #SCK035)
For parts to fit the SciCan STATIM 2000
Water Filter (In-Line) SCF034
or Pump Mesh Filters
Other Routine –
Maintenance
PART RPI PART # STERILIZER PM TIP
Debris is the #1 reason for pump failure, so before replacing this filter, be
sure to check the reservoir for dirt and debris. Inspect and clean or replace
the Reservoir Cap & Filter (RPI Part #SCK059). Drain, clean and flush the reser-
voir. Refill reservoir with distilled water after filter replacement.
If your autoclave comes equipped with an In-Line Water Filter, cut the cable
ties and remove the old filter. Note the flow arrow direction on the filter body
(flow direction is from the reservoir to the pump). Install the new filter using
the new cable ties included. Be sure to prime the pump in order to purge
trapped air before running a complete cycle. If your machine does not have
an In-Line Water Filter, then be sure to clean the pump mesh filters (located
inside the pump’s input and output port fittings). See page 6 and 7 - Figure
2.3 A and B.
The OEM recommends replacing the Water Filter (In-Line) every 6 months or
500 cycles – whichever comes first.
If the autoclave you’re working on has the mesh filters, we recommend you
upgrade to the Drain Kit with Pump Filter (RPI Part #SCK037), making future
service much easier and providing a much better filtering system to protect
the pump (along with providing a quick & easy way to drain the reservoir
through the front of the autoclave).
Apply Sci-Dry™ (RPI Part #SCA054) every 10 cycles and after cleaning. Coat
all internal surfaces of cassette. Sci-Dry™ minimizes spotting and allows in-
struments to dry more efficiently.
The OEM recommends the replacement of the Check Valve Kit (RPI Part
#SCK011) and the steam generator/boiler’s Safety Valve (70PSI) (RPI Part
#SCV004) every 2 years, and the use of the Reservoir Cap & Filter (RPI Part
#SCK059) on all versions of the STATIM autoclave.
To aid in proper cleaning and maintenance of autoclaves, RPI also offers a
Cleaning Kit (RPI Part #RPK791) which includes a variety of brushes, sponges
and cleaning pads assembled in a convenient carrying case.
Sterilizer PM Kit (RPI Part #SCK035)