2.3 - PUMP TESTER BOTTLE
(RPI PART #SCK024)
The RPI bottle has a “swivel fitting” on the lid that allows
you to control the position and balance of the bottle. In ad-
dition, RPI includes 8” of tubing already connected to the
swivel fitting which allows you to move the bottle a safe
distance away from the autoclave during testing.
The OEM now recommends the use of the Reservoir Cap &
Filter (RPI Part #SCK059) along with the In-Line Water Filter
(RPI Part #SCF034) on all units. The mesh filters should be re-
moved and the autoclave should be upgraded to the In-Line
Water Filter (Drain Kit with Pump Filter RPI Part #SCK037).
INSTRUCTIONS
STEP #1 - TO BEGIN: Disconnect the pump outlet meter-
ing tube (see Figure 2.3A) from the steam generator/
boiler by using a 3/8” wrench on the compression nut
while supporting the steam generator/boiler fitting with a
FI
LL LI
N
E
M
I
NIMUM
O
I
Flow
290
7/16” wrench. Connect the metering tube to the Pump
Tester Bottle male connector and tighten.
STEP #2 - PRIME PUMP: Fill reservoir with distilled water
only and activate the pump for 2-3 seconds to purge any
trapped air from the tubing (until water passes through
the pump) (see page 5, 2.2 - Start Switch). Empty the
bottle of any water that may have entered it during the purge
cycle. The bottle must be completely empty before proceed-
ing with the test.
STEP #3 - TEST CYCLE: Before activating the pump for
the Test Cycle, have a stopwatch or another timing device
available for use during the Test Cycle. Activate the pump
and note the amount of time that it takes for the water
level in the bottle to reach the MINIMUM FILL LINE.
If the water level reaches the MINIMUM FILL LINE within
25 seconds, pump testing is complete and no further
steps are required.
If the water level does not reach the MINIMUM FILL LINE
within the specified time, proceed to the next step.
STEP #4 - FILTERS:
If the autoclave has been upgraded with Drain Kit
with Pump Filter (RPI Part #SCK037), be sure to re-
move the mesh filter from both the inlet and outlet fit-
tings (if they have not already been removed). The ULKA
brand pump does not have the mesh filters. Replace the
Water Filter (In-Line) (RPI Part #SCF034). Repeat Step
#2 - Prime Pump, then if the water level still does not
reach the MINIMUM FILL LINE within 25 seconds, re-
place the pump.
If the unit does not have the Drain Kit with Pump
Filter (RPI Part #SCK037) installed, then clamp off the
pump inlet tubing from the reservoir using hemostats or
a tubing clamp (Note: It is important to clamp the tubing
so that the reservoir does not drain). Disconnect the me-
tering tube from the pump. Remove the filtered elbow
outlet fitting and the filtered barb inlet fitting from the
pump. (Caution: Use a wrench on the pump body while
290
Flow
Flow
FIGURE 2.3B
TIP - If the unit does not have the
Drain Kit with Pump Filter (RPI
Part #SCK037) and the Reservoir
Cap & Filter (RPI Part #SCK059),
now is a good time to upgrade. If
time does not permit an upgrade,
remove and clean mesh filters
inside inlet and outlet fittings as
described above and anticipate
an upgrade at the next service.
removing filtered fittings). Clean the mesh filter of both
the inlet and the outlet fittings by placing the fittings
under running water, then blow both ends out with an air
hose. Reinstall the filtered fittings using Teflon® tape
(see Figure 2.3B). Be sure to prime the pump in order
to purge trapped air before running a complete cycle. If
the water level still does not reach the MINIMUM FILL
LINE within 25 seconds, replace the pump (see
Tip
in
Figure 2.3B). For additional information regarding the
pump see page 43 - Figure 9.6.
RELATED RPI PARTS
:
• Drain Kit with Pump Filter (RPI Part #SCK037)
• Metering Tube (RPI Part #SCT053)
• Reservoir Cap & Filter (RPI Part #SCK059)
• Teflon
®
Tape (RPI Part #RPT579)
Water Filter (In-Line) (RPI Part #SCF034)
Drain Kit with Pump Filter
(RPI Part #SCK037);
Filter only (RPI Part #SCF034)
Metering Tube
(RPI Part #SCT053)
Reservoir Cap & Filter
(RPI Part #SCK059)
FIGURE 2.3A
TIP - The OEM now recommends the use
of the Reservoir Cap & Filter (RPI Part
#SCK059) along with the In-Line Water
Filter on all units. The mesh filters should
be removed and the autoclave should be
upgraded to the In-Line Water Filter (Drain
Kit with Pump Filter RPI Part #SCK037).
Bottle Male Connector
Bottle Inlet Tube (8”)
Pump Tester Bottle
(RPI Part #SCK024)
Inlet Fitting
Outlet Fitting
Metering Tube (RPI Part #SCT053)
to Steam Generator/Boiler
Mesh Filter
Reservoir